Picoway super picosecond
one,Introduction
The picosecond laser is a laser whose pulse duration (pulse width) emitted by each laser reaches the picosecond level; the most cutting-edge technology is used to treat pigmented diseases. Through fast and powerful energy, the pigment in the skin is more finely crushed, and then the principle of removing tattoos and stains through the principle of lymphatic drainage. Therefore, the treatment process is shortened from the original 10 times to 2 to 3 times, and the dermis is not harmed, the possibility of side effects is reduced, and the success rate of the operation is improved.
two,Treatment range
Remove tattoos, eyeliner, lip liner Epidermal and dermal pigments Acquired Ota mole-like spot (stain on both sides) freckle Black spot... | Pigmentation after inflammation Age spots Sunburn/simple spots Coffee spot
|
three,Advantages and characteristics
1. High-energy & rapid treatment: short-term high-energy radical treatment of stains (tattoos, epidermal spots, dermal spots)
2. High-end performance: Picosecond-level high-speed crushing of large pigment cell tissues into fine debris
3. Comfort and safety: It can effectively and safely treat various pigmented diseases and stubborn spots, because picosecond laser treatment can reduce the damage to the skin by accurately positioning the target tissue to achieve the effect of removing spots.
4. The skin will not be scalded during treatment: because the picosecond laser has only half the energy of traditional lasers, the thermal damage to the skin tissue is also reduced by half
5. There will be no anti-black problem: the picosecond laser energy is instantly focused and penetrates to the skin surface, accelerates the decomposition and metabolism of melanin particles, and is not easy to remain on the skin, greatly reducing postoperative redness and anti-black phenomenon.
6. Honeycomb-type instant lens: Causes the epidermal vacuolation effect, can protect the epidermis from wounds, achieve the mechanism of initiating tissue repair, and provide more diverse treatments.
Fourth, the principle of treatment
Conversion relationship of time unit:
1 second = 1000 milliseconds (Millisecond)
1 millisecond = 1000 microseconds (Microsecond)
1 microsecond = 1000 nanoseconds (Nanosecond)
1 nanosecond = 1000 picoseconds (Picosecond)
As the name implies, a picosecond laser is a laser whose pulse duration (pulse width) emitted by each laser reaches the picosecond level.
According to the principle of selective photopyrolysis, the shorter the action time of the laser, the less likely it is for the laser energy absorbed in the target tissue to diffuse to the surrounding tissues, and the energy is confined to the target to be treated to a degree, protecting the surroundings Normal tissues, and thus the more selective the treatment. The pulse width of the picosecond laser is only one hundredth of that of the traditional Q-switched laser. Under this ultra-short pulse width, the light energy is too short to be converted into heat energy, and there is almost no photothermal effect. After being absorbed by the target, its volume expands rapidly. The photomechanical effect is torn into fragments by blasting, and the selectivity is stronger, and the pigmented skin lesions can be treated with a shorter number of times to produce a stronger curative effect. In one sentence, “the picosecond laser breaks the pigment particles more thoroughly, while causing less damage to the surrounding tissues.”
If we compare the original pigment particles to rocks, the traditional Q-switched laser can break these rocks into the size of pebbles, and the picosecond laser can break them into fine sand, so that the absorption efficiency of pigment fragments is improved. Greatly improved.
Five, accessories list
Accessories box Korean torsion spring light guidearm | Treatment handle | Pedal |
Honeycomb head | Add water funnel | Handle hanger |
Eye masks & glasses | Blindfold | power cable |
Hex screwdriver |
Six, technical parameters
Technical Data Sheet | |
Laser type | ND:YAG |
Laser wavelength | 532nm, 1064nm (755nm optional) |
Laser output energy | Single pulse 2.00j |
power supply | 220V/110V |
Highest energy | 6J/cm |
working frequency | 1-10HZ |
Pulse width | 2000ps |
Total power | 3000W |
cooling system | Air-cooled, water-cooled |
Spot diameter | 2-10mm |
Net amount | 140KG |
Package dimensions |
Seven, safety matters
1. Electrical safety
The laser uses a single-phase 220V/5A or 110V/10A power supply, and the maximum output current is not more than 3A.
The power supply connected to the laser must use a single-phase three-wire power socket that meets the national standard and is well grounded, with a specification of at least 5A.
There is high voltage in the memory when the laser is running. After the power is turned off, the high voltage devices may still carry residual voltage-open the case at will, there may be a high voltage electric shock hazard!
2. Safe handling
The laser is a large-scale equipment and is not suitable for users to patrol and move. The handle shell of the laser machine is equipped with a precise laser optical system. The shock-absorbing packaging provided by the original manufacturer must be used during the handling process, otherwise it may cause the laser system to be out of adjustment and affect the laser output energy!
The drain hole must be plugged tightly before moving and transporting
3. System safety device
In order to ensure safe use, the laser is designed with the following safety protection devices:
Stop switch: convenient for emergency shutdown in case of accidents;
Standby/ready control: The laser output is prohibited in the standby state, which is the safe state of the system.
4. [Declaration]: Low temperature area below 0℃
If the instrument is not used for a long time or needs to be used in a different place, the water inside (handle and host) must be drained.
Try not to place the instrument in an environment lower than 0 ℃ for more than 3 hours. Research has found that water below 0 ℃ will "explode" when moving. Specifically, it turns into liquid water below 0 ℃. There will be rapid freezing in the process. After the liquid water turns into solid ice, its volume increases rapidly, causing certain damage to the water pipes, the internal waterways of the cooling fan, and the laser waterways.
After receiving the instrument, the customer needs to stand for 24 hours in an environment with an indoor temperature of 23-26°C before it can be used. If the operation is not performed according to the requirements, due to the customer's own reasons, the instrument is damaged, the company will only provide paid maintenance services. Due to the influence of temperature, we hope that customers can pay attention to it. And follow the requirements and carry out routine maintenance.
8. Operating instructions
- Instrument installation:
1.1 Installation of light guide arm:Take out the light guide arm and put it directly into the light guide arm connection port of the host, and then rotate the nut to fix it;
1.2 Installation of handle hanger:Fix the handle hanger on the light guide arm with screws;
①Hanging jack
②Screw fixing
③Complete
1.3 Installation of treatment handle:Follow the arrow to align the socket and insert and fasten
1.4 Add water:Insert the drain pipe into the overflow hole on the left side of the main unit, insert the funnel into the water filling hole, and add distilled water through the funnel until water flows out of the overflow hole and it is full;
Drain the water:Insert the drain pipe into the drain hole on the right side of the main unit, insert the funnel into the water filling hole, and the water will be discharged directly through the drain pipe;
Note: When adding water, place a container at the corresponding position of the overflow to catch the water
It is not allowed to turn on without water at any time
Use pure water or distilled water as coolant
Replace water with antifreeze when the temperature is below 0°
Change the water once a week, and empty the water when it needs to be transported
1.5 Foot switch and power connection:The foot switch is used to control the laser output. When the foot switch is pressed down, the shutter of the shutter motor opens and the laser is output from the end of the articulated arm.
2. Screen introduction
2.1 Boot:Turn on the power, turn to the right to pop up the power buttonTo boot;
Button key
Boot interface
2.2 Screen introduction:
Repeatedly adjust the energy level to 0.1-6J/cm² in the standby state;
Record the number of current operations;
Adjust the red light indicator;
Energy display
Frequency display
Spot size display;
(Remark:
The treatment handle can be rotated to adjust the spot size by pushing it up slightly2-10)
Application wavelength, optional 1064nm&755nm&532nm
Nine, parameter settings
Parameter setting reference table:(The actual operation situation is adjusted according to the situation reflected in the treatment of the patient) | ||||
Light spot | Energy J/cm² | Wavelength nm | Remark | |
Whitening and removing yellow | 7 | 0.9-1.2 | 1064/755 | Operation until the skin is reddish |
Epidermal spots | 4-5 | 1.0-1.2 | 1064/755 | Slight oozing blood after operation until the spots turn white |
Dermal spots | 4 | 0.9-1.2 | 1064 | Operate to reddish or Sha |
chloasma | 6 | 0.7-1.0 | 1064 | The operation becomes whitish or reddish |
Mole | 4 | 1.0-1.3 | 1064 | Operation until bleeding or whitening |
Eyebrow tattoo, tattoo, lip line, eyeliner | 6 | 1.3-1.5 | 1064 | New style tattoo, black and blue, the tattoo turns white after the operation |
6 | 0.9-1.0 | 1064 | Old-fashioned tattoo, black and blue, until the tattoo turns white | |
5 | 0.9-1.0 | 532 | Wash the color until the tattoo turns white | |
Blackhead | 5 | 1.1-1.3 | 1064 | Knock off the little black head |
X. Effect comparison chart
11. Operation precautions and contraindications
1,Precautions
①It is normal to feel slight tingling during treatment;
②Can not be applied to pregnant women;
③It cannot be used for bleeding, thrombosis, arteriosclerosis, acute inflammation, and suppuration of joints and soft tissues;
④It is forbidden to store near direct light and heating appliances, especially in the summer, choose a good place;
⑤It is easy to get up and stay away from places with water or humidity (sauna rooms, bathing halls, etc.), which will cause damage;
⑥ It is forbidden to be impacted by external force to avoid damage, especially when using the handle, be careful not to fall on the ground;
⑦It is forbidden to use chemical drugs to wipe the instrument (alcohol, benzene, senna water, etc.), which will cause damage;
⑧Do not use the instrument continuously for a long time, it will cause damage;
⑨Do not let children touch the instrument;
10 This instrument should avoid sensitive parts such as eyes, larynx, sexual organs, etc.;
11Do not open the inside of the instrument randomly.
2、It is forbidden to use this product for physical therapy in the following situations
1Patients with edema or edema, patients with hypertension;
2 Those who have severe convulsions or irritability throughout the body;
3 Heart disease patients (persons with cardiac pacemakers, etc.);
4 People with skin allergies, skinny or loss of elasticity, skin breakage, ulceration;
5 Severe blood loss, bleeding disorders, bleeding tendency or coagulation difficulties;
6 People who do not know how to operate the instrument;
7 Those who have undergone facial surgery or deep exfoliation, but have not yet reached the recovery period;
8 People who have received radiotherapy for less than 3 months;
9 People who have skin cancer at the physical treatment site;
10 People who have skin disease or inflammation, and people with purulent acne;
11 It is forbidden to use it at the early stage after the operation or in the traumatic part;
After-sales instructions
- From the day the instrument leaves the factory (subject to the bar code of the machine), Party A will provide Party B with a one-year free maintenance and lifetime maintenance after-sales service for the host (excluding the handle).
- From the day the instrument leaves the factory (according to the machine's factory barcode), the handles of all models are guaranteed for three months free of charge, except for human factors. (Human factors refer to unconventional instrument operations, such as not adding water, breaking, etc.)
- The laser sighting head and the E-light insert are not covered by the warranty. For the consumables that have a fixed number of points for the treatment head, we guarantee half of the points.
- Within 7 days since the instrument leaves the factory (15 days abroad), if the machine has quality problems (non-human and force majeure), our company will provide a free replacement service for the new machine. If the customer sends back according to our company's designated freight, our company will bear the two-way freight.
- From 15 days to one year from the factory, if the machine has quality problems (non-human and force majeure), our company will provide free maintenance services and free replacement parts services. If the whole machine is not returned to the factory for repair, you can do it yourself under the guidance of our engineers Remove the accessories and return to the factory. After the accessories are repaired, we will guide the installation. If you need to provide accessories first due to long distances, urgent time, etc., we will charge the accessories first, and the damaged accessories will be returned to the factory within one month for a full refund, and those that cannot be returned in time will be regarded as accessories shipped (for foreign customers within three months ). The original packaging must be kept when the whole machine is returned to the factory. We are not responsible for the damage caused by packaging problems. The customer and our company are each responsible for the one-way delivery cost of the accessories or the complete machine. If engineers need on-site service, the customer provides round-trip travel expenses (except Guangzhou);
- If the equipment has been out of the factory for more than one year, if there is a quality problem with the machine, Party A will provide Party B with a paid warranty, and only charge parts replacement fees (the replacement parts warranty period is 3 months), and the travel expenses of the engineers. If you need to return to the factory for repair, you will be charged two-way freight and parts replacement fees.
Remarks: All the products produced by our company are precision instruments and cannot be transported frequently. Due to the quality of the instrument caused by the handling, we can provide a warranty, but the responsibility is not on our side. The original packaging is very important for the future upgrade and repair of the instrument, and it is recommended that all customers keep it. The factory barcode is the only certificate of warranty, and the warranty service is not provided if the barcode is torn up. The company reserves the right of final interpretation and modification of the after-sales description.